Control system: TESAMODUL block. Specifications:
Parameter | Value | Units |
Operating range: | ||
Minimum nominal length of the controlled end measures | 0,5 | мм |
Maximum nominal length of the controlled end measures | 100 | мм |
Dimensions of the cross-section of the controlled end measure | 9 x 30 or 9 x 35 | mm |
Measuring force: | ||
upper sensor | 1 | Н |
lower sensor | 0,63 | Н |
Dimensions: > | ||
Lenghth | 180 | mm |
Width | 200 | mm |
Height | 550 | mm |
Uncertainty of measurement: | ||
Length measurement error | ± (0,05+0,5xL) | mkm, L – in meters |
Repeatability (at constant temperature) | ± 0,015 | mkm |
Power supply: | ||
Electrical network | 220 | V |
Frequency | 50-60 (± 2%) | Hz |
Requirements for environmental conditions: | ||
Operating temperature | +19,5…+20,5 | °С |
Features
Measuring system for monitoring the end measures contains: a measuring stand, a special measuring table with 6 carbide inserts for positioning the end measure; registration system of two inductive sensors; electric vacuum pump for pneumatic discharge of measuring tips; block TESAMODUL for calculation and output of the measured deviation parameters with data transfer to a personal computer.
Specification of device supply
Accessories | Type |
Measuring system basic kit of increased accuracy (type 59.30003) for work with a personal computer consists of: | |
basic mechanical unit with registration system | 16.10401 |
Electric vacuum pump with remote control | 32.60433 |
pneumatic suction cup for moving end measures up to 10 mm | 16.60011 |
electronic unit TESAMODUL models 372/404/708 | 41.30105 |
Built-in card for digital value output, including interface RS232 | 41.60079 |
connecting cable between the card and the interface unit | 47.61050 |
connecting cable between the interface unit and the vacuum pump | 47.61005 |
remote pedal for starting measurement and data transfer to a computer | 47.68000 |