ISO 9001:2015
Modern means of measuring equipment
 

Measuring instrument for monitoring of end measures TESA UPC

Production:
Hexagon MI TESA (Switzerland)

The device serves to calibrate (dimension control) plane-parallel end measures with a rectangular cross-section in the range of nominal sizes from 0.5 to 100 mm. A method of comparative measurement using control end measures is used.

Measuring instrument for monitoring of end measures TESA UPC
Measuring instrument for monitoring of end measures TESA UPC

Control system: TESAMODUL block. Specifications:

Parameter Value Units
Operating range:
Minimum nominal length of the controlled end measures 0,5 мм
Maximum nominal length of the controlled end measures 100 мм
Dimensions of the cross-section of the controlled end measure 9 x 30 or 9 x 35 mm
Measuring force:
upper sensor 1 Н
lower sensor 0,63 Н
Dimensions:
Lenghth 180 mm
Width 200 mm
Height 550 mm
Uncertainty of measurement:
Length measurement error ± (0,05+0,5xL) mkm, L – in meters
Repeatability (at constant temperature) ± 0,015 mkm
Power supply:
Electrical network 220 V
Frequency 50-60 (± 2%) Hz
Requirements for environmental conditions:

Operating temperature +19,5…+20,5 °С

Features
Measuring system for monitoring the end measures contains: a measuring stand, a special measuring table with 6 carbide inserts for positioning the end measure; registration system of two inductive sensors; electric vacuum pump for pneumatic discharge of measuring tips; block TESAMODUL for calculation and output of the measured deviation parameters with data transfer to a personal computer.

Specification of device supply

Accessories Type
Measuring system basic kit of increased accuracy (type 59.30003) for work with a personal computer consists of:
basic mechanical unit with registration system 16.10401
Electric vacuum pump with remote control 32.60433
pneumatic suction cup for moving end measures up to 10 mm 16.60011
electronic unit TESAMODUL models 372/404/708 41.30105
Built-in card for digital value output, including interface RS232 41.60079
connecting cable between the card and the interface unit 47.61050
connecting cable between the interface unit and the vacuum pump 47.61005
remote pedal for starting measurement and data transfer to a computer 47.68000
execution of the instrument with increased accuracy (and with a calibration certificate) 16.90021

Additional information